15x15 Carbon Monoxide Sensor 1000 ppm Package 110-102 - 3SP_CO_1000 P

• Measurement Range: 0 to 1,000 ppm
• Resolution: < 100 ppb (instrumentation dependent)
• Repeatability: < ± 2 % of reading
• Response Time T(90): < 30 seconds
• Sensitivity: 4.75 ± 2.75 nA/ppm
• Operating Temperature: -30 to 55 °C
• ROHS Compliant

  • Model Number: 3SP_CO_1000 P
  • Data Sheet: Date Sheet File
Quantity:
- +
  • 15x15 Carbon Monoxide Sensor 1000 ppm Package 110-102
  • 15x15 Carbon Monoxide Sensor 1000 ppm Package 110-102
Mouse over image to zoom

Product Specification

3SP_CO_1000 P
15x15 CO Sensor 1000 ppm Package 110-102


3SP_CO_1000 P Carbon Monoxide Sensor CO Sensor Description

SPEC Sensors' Screen Printed ElectroChemical sensor technology (SPEC Sensor) revolutionizes the current state of the art, enabling new applications in consumer and industrial safety monitoring. SPEC’s printed sensors offer the performance of the best quality electrochemical sensors at a fraction of the price. SPEC' s printed sensors are also ultra-thin, offering easy integration into wireless, portable, and networked solutions. These sensors are ideal for health, environmental, industrial and residential monitoring, because of their high performance, low cost and small size.


3SP_CO_1000 P Carbon Monoxide Sensor CO Sensor Benefits

• Small Size with Low Profile (20x20x3.8 mm)
• Long Life (10 years expected life)
• Fast Response (< 15 seconds)
• Robust (passes 5000 ppm overload)
• Low Power (0 mW @ 0 mV bias)
• Individually Calibrated (NIST Traceable)
• ROHS Compliant


3SP_CO_1000 P Carbon Monoxide Sensor CO Sensor Applications

• Residential and Commercial CO Monitoring
• Industrial CO Monitors
• Ventilation Control
• RV and Marine CO Monitoring
• Indoor Air Quality
• Outdoor Air Quality


3SP_CO_1000 P Carbon Monoxide Sensor CO Sensor Specifications

Measurement Range

 0 to 1,000 ppm

Detection Limit

 0.5 ppm

Resolution

 < 100 ppb (instrumentation dependent)

Repeatability

 < ± 2 % of reading

Response Time - T(90)

 < 30 seconds (15 seconds typical)

Sensitivity

 4.75 ± 2.75 nA/ppm

Overload

 Passes EN20291-1 Sec. 5.3.6 5,000 ppm overload

Expected Operating Life

 > 5 years (10 years @ 23 ± 3 °C; 40 ± 10% RH)

Operating Temperature Range

 -30 to 55 °C (-20 to 40 °C continuous recommended)

Operating Humidity Range - non-condensing

 15 to 95% recommended continuous

 0 to >95% RH - intermittent

Operating Bias

 0 to 5 mV

Power Consumption

 10 to 50 uW (circuit & ambient CO dependent)


3SP_CO_1000 P Carbon Monoxide Sensor CO Sensor Cross Sensitivity
Most chemical sensors exhibit some cross-sensitivity to other gases. The following table lists the relative response of common potential interfering gases, and the concentration at which the data was gathered.

Gas/Vapor

Concentration

Typical Response PPM CO

Carbon Dioxide

5,000 ppm

< 1

Hydrogen

100 ppm

17

Methane

3,000 ppm

< 1

Ammonia

100 ppm

< 1

Nitrogen Dioxide

10 ppm

< 1

Hydrogen Sulfide

25 ppm

< 1

Carbon Monoxide

400 ppm

400

Ozone

5 ppm

< 1

Sulfur Dioxide

20 ppm

< 1

Chlorine

10 ppm

< 1

n-Heptane

500 ppm

< 1

Toluene

200 ppm

< 1

Isopropyl Alcohol

200 ppm

1.3

Acetone

200 ppm

< 1


If you cannot find what you want, you can entrust ISweek to source for you. Just click: