Flange Mount Zirconia Oxygen Sensor - KGZ-10 Series

• Oxygen pressure range: 2 mbar to 3 bar
• Operational temperature: 700°C [1292°F]
• Sensitivity: 1.05 ms/mbar​
• Accuracy: <5 mbar​
• Response time: <4 s
• Warm up time: <100 s​
• Gas flow rate​: 0 m/s to 10 m/s​

  • Model Number: KGZ-10 Series
  • Data Sheet: Date Sheet File
Quantity:
- +
  • Flange Mount Zirconia Oxygen Sensor
  • Flange Mount Zirconia Oxygen Sensor
Mouse over image to zoom

Product Specification

KGZ-10 Series Zirconia Oxygen Sensor Description
The KGZ-10 is a high-accuracy, zirconia-based oxygen sensor. The ability of the sensor to operate without any reference gas makes it ideal for potential applications such as combustion control or aircraft oxygen generation systems.

The sensor employs two ZrO2 discs with a small hermetically sealed chamber in between. One of the discs functions as a reversible oxygen pump, which is used to successively fill and empty the chamber. The second disc measures the ratio of the partial pressure difference and generates a corresponding sense voltage.

A heat element produces the 700 °C [1292 °F] required for the ZrO2 to operate as an O2 pump. The time taken for the pump to achieve specific minimum and maximum pressures within this chamber is a measure of the partial oxygen pressure of the environment.

The KGZ-10 has to be operated by an electronic measuring circuit that controls the sensor operation and signal processing. This circuit can either be incorporated into the customer s own electronics, or be purchased as a separate interface board from Honeywell.

KGZ-10 Series Zirconia Oxygen Sensor Features
• High-accuracy measurement
• Low power consumption
• No reference gas required
• Linear output signal
• No need for temperature stabilization
• Function testing and calibration in ambient air
• Long life

KGZ-10 Series Zirconia Oxygen Sensor Potential Applications
• Boiler combustion controls
• Aircraft oxygen generation systems

KGZ-10 Series Zirconia Oxygen Sensor Technical Specifications

Characteristic
Parameter
Sensor voltage levels (recommended)
45-64-85 mV
Pump current (recommended)
40 uA
Heater supply
4.0 V (1.7 A)
Heater supply (stand by)
2.0 V
Pump resistance at 700°C [1292°F]:
    dc
    ac

1 kOhm, typ.
1 kHz 120 Ohm, typ.
Oxygen pressure range
2 mbar to 3 bar
Operational temperature
700°C [1292°F]
Stand by temperature
500°C [932°F], typ.
Sensitivity 
1.05 ms/mbar
Accuracy
<5 mbar
Response time
<4 s
Warm up time
<100 s
Warm up time (from stand by)
<20 s
Permissible gas temperature
-100°C/250°C [-148°F/482°F]                    
Gas flow rate
0 m/s to 10 m/s
Repetitive permissible acceleration
5 g
Incidental permissible acceleration
30 g

KGZ-10 Series Zirconia Oxygen Sensor Order Guide
Catalog Listing
Description
KGZ-10
Oxygen sensor, flange mount with tabs
KGZ-10SP
Oxygen sensor, flange mount with pins

If you cannot find what you want, you can entrust ISweek to source for you. Just click: