Chlorine Sensor (CL2 Sensor) - CL2-B1

• Range: 20 ppm limit of performance warranty
• Sensitivity: -600 to -1150 nA/ppm in 10ppm Cl2
• Response time t90: < 60 (s) from zero to 10ppm Cl2
• Resolution: < 0.02 RMS noise (ppm equivalent)
• Operating life: > 24 months
• Temperature range: -20 to 50​°C
• Pressure range: 80 to 120 kPa
• Humidity range: 15 to 90​ % rh continuous
• Weight: < 13 g

  • Model Number: CL2-B1
  • Data Sheet: Date Sheet File
Quantity:
- +
  • Chlorine Sensor (CL2 Sensor)
  • Chlorine Sensor (CL2 Sensor)
Mouse over image to zoom

Product Specification

CL2-B1 Chlorine Sensor (CL2 Sensor) Description
Alphasense Chlorine gas sensors operate using proven fuel cell technology.

CL2-B1 Chlorine Sensor size:
32mmdiameter package, the best choice for fixed site application

The Cl2 range provides OEMs with reliable sensors for use in a number of high volume applications. Strong signal levels combined with low zero current allows resolution to 20ppb and an operating range from +0 to 20ppm. The sensors are suitable for inclusion in fixed installation sensing heads and portable safety instruments.

These sensors have improved transient response to humidity changes compared to competitors’ equivalent sensors.

The Cl2 sensors offer our electrolyte leak-free guarantee and reliable long-term detection performance. The leakproof housing is moulded with a colour coded top for ease of identification.

CL2-B1 Chlorine Sensor (CL2 Sensor) PERFORMANCE
Sensitivity
nA/ppm in 10ppm Cl2
-600 to -1150
Response time
t90 (s) from zero to 10ppm Cl2 (33Ω load resistor)
< 60
Zero current
ppm equivalent in zero air
± 0.2
Resolution
RMS noise (ppm equivalent) (33Ω load resistor)
< 0.02
Range
ppm limit of performance warranty
20
Linearity
ppm error at full scale, linear at zero and 10ppm Cl2
< ± 0.2 
Overgas limit
maximum ppm for stable response to gas pulse
60



CL2-B1 Chlorine Sensor (CL2 Sensor) LIFETIME
Zero drift
ppm equivalent change/year in lab air
< 0.03
Sensitivity drift
% change/year in lab air, monthly test
< 6
Operating life
months until 80% original signal (24 month warranted)
> 24



CL2-B1 Chlorine Sensor (CL2 Sensor) ENVIRONMENTAL
Sensitivity @ -20°C
% (output @ -20°C/output @ 20°C) @ 10ppm
70 to 90
Sensitivity @ 50°C
% (output @ 50°C/output @ 20°C) @ 10ppm
90 to 105
Zero @ -20°C
ppm equivalent change from 20°C
< 0 to 0.1
Zero @ 50°C
ppm equivalent change from 20°C
< 0 to -1



CL2-B1 Chlorine Sensor (CL2 Sensor) CROSS SENSITIVITY
H2S   sensitivity
% measured gas @ 20ppm H2S
-100
NO2   sensitivity
% measured gas @ 10ppm NO2
100
NO    sensitivity
% measured gas @ 50ppm NO
< 0.5
SO2   sensitivity
% measured gas @ 20ppm SO2
< -2
CO     sensitivity
% measured gas @ 400ppm CO
< 0.1
H2     sensitivity
% measured gas @ 400ppm H2
< 0.1
C2H4 sensitivity
% measured gas @ 400ppm C2H4
< 0.1
NH3   sensitivity
% measured gas @ 20ppm NH3
< 0.1
CO2  sensitivity
% measured gas @ 5% (Vol) CO2
0



CL2-B1 Chlorine Sensor (CL2 Sensor) KEY SPECIFICATIONS
Temperature range
°C
-20 to 50
Pressure range
kPa
80 to 120
Humidity range
% rh continuous (see note below)
15 to 90
Storage period 
months @ 3 to 20°C (stored in sealed pot)
6
Load resistor
Ω (for optimum performance)
33
Weight
g
< 13


Matched Transmitter board for CL2-B1 CL2 Sensor

4-20mA Analogue Transmitter Board
  • Factory calibration, custom-set for immediate use
  • Wide operating voltage range
  • Low RFI susceptibility
Alphasense 4-20 mA analogue transmitter boards provide a cost effective way for Original Equipment Manufacturers to engineer transmitter heads for fixed installation systems.


4-20mA Digital Transmitter Board
   • Factory calibration, custom-set for immediate use
   • Digital zero, sensitivity and temperature compensation
   • Digital and 4-20mA output
   • Biased or unbiased operation
   • Calibration and range change

If you cannot find what you want, you can entrust ISweek to source for you. Just click:

Sourcing Service