Submersible Water Quality Sensor Turbidity Sensor - WQ730

• Range: 0-50 NTU and 0-1000 NTU
• Accuracy: +/- 1% of full scale​
• Meter Resolution: 12 bit
• Output: 4-20mA
• Operating Voltage: 10-36 VDC @ 40 MS
• Current Draw: 30 mA plus sensor output
• Warm Up Time: 5 seconds minimum
• Ope​rating Temperature: 14 to 122°F (-10 to +50°C)
• Maximum Pressure: 30 psi

  • Model Number: WQ730
  • Data Sheet: Date Sheet File
Quantity:
- +
  • Submersible Water Quality Sensor Turbidity Sensor
  • Submersible Water Quality Sensor Turbidity Sensor
Mouse over image to zoom

Product Specification

WQ730 Turbidity Sensor Description
WQ730 Turbidity Sensor is a highly accurate submersible instrument for in-situ environmental or process monitoring. Applications for the WQ730 include: water quality testing and management, river monitoring, stream measurement, reservoir water quality testing, groundwater testing, water and wastewater treatment, and effluent and industrial control.

WQ730 Submersible Water Quality Sensor Turbidity Sensor Features
• 4-20 mA output
• Marine grade cable with strain relief
• In-situ turbidity measurement
• Simple and convenient to use
• Rugged stainless steel and Delrin sensor housing

WQ730 Turbidity Sensor Specifications

Range 0-50 NTU and 0-1000 NTU
Accuracy
+/- 1% of full scale
Meter Resolution
12 bit
Output
4-20mA
Method
Nephelometer with correction
Operating Voltage
10-36 VDC @ 40 MS
Current Draw
30 mA plus sensor output
Warm Up Time
5 seconds minimum
Operating Temperature
14 to 122°F (-10 to +50°C)
Materials
306 stainless steel, delrin, polyether jacketed cable
Maximum Pressure
30 psi
Light Source
Infrared LED, (880nm)
Cable Length
25 ft standard (optional to 500 ft)
Size
Body= 1 1/2 x 8.5 inches (3.8 x 21.6 cm) (Dia x Length)
Weight
1lb (454 g) (Sensor)

If you cannot find what you want, you can entrust ISweek to source for you. Just click:

Sourcing Service