OEM Photoionization Sensor (PID Sensor) - 2 ppm - PID-2

• PID-200: Blue
• Range: 2 ppm
• MDQ: 0.5 ppb
• T90 Response Factor: ≤ 4 sec (typical)
• Supply Voltage: 3.2 V to 5.5 V
• Current: 24 mA to 36 mA
• Temperature Range: -20 °C to 60 °C (-4 °F to 140 °F)
• Relative Humidity Range: 0 to 90% non-condensing

  • Model Number: PID-2
  • Data Sheet: Date Sheet File
Quantity:
- +
  • OEM Photoionization Sensor (PID Sensor) - 2 ppm
Mouse over image to zoom

Product Specification

piD-TECH eVx OEM Photoionization Sensor (PID Sensor) Description
The piD-TECH® eVx family of photoionization detectors offers the latest in technology and upgrades without compromising the award winning features that make the piD-TECH line of sensors the best choice for OEM Manufacturers. Providing better value, design flexibility and lower detection limits. These improvements include greater stability over temperature and humidity changes, integral shielding, improved response time and minimum detection limits down to 0.5 ppb.

These detectors are designed intrinsically safe and with a 4P platform for ease of installation in both portable and fixed gas detection units. For use in a myriad of VOC (volatile organic compound) monitoring applications including environmental and safety applications in the industrial, commercial or residential markets. 


piD-TECH eVx OEM Photoionization Sensor (PID Sensor) Applications
• Industrial Hygiene & Safety Monitoring
• Confined Space Entry
• Soil Contamination and Remediation
• Hazmat Sites & Spills
• Low Concentration Leak Detection
• Indoor Air Quality
• Fenceline Monitoring
• EPA Method 21

piD-TECH eVx OEM Photoionization Sensor (PID Sensor) Features & Benefits
• Designed for use in intrinsically safe circuits: UL, CAN/CSA, ATEX, IECEx certified
• Integral shielding
• 4P cell platform compatible
• Lamp re-Ignition circuit for fast, repeatable and dependable startup
• Reliable long lamp life: guaranteed 6,000 continuous hours
• Internal, input voltage regulation for greater signal stability
• Dual filter technology protects from aerosols and particulates
• Negligible temperature response
• Guard electrode curtails humidity effects
• Rapid response & greater baseline stability
• Easily cleaned and field serviced
• Out of warranty Sensor Exchange Program

PID-2 OEM Photoionization Sensor (PID Sensor) Specifications


Blue
Part Number
045-014
Range**
2 ppm
MDQ**
0.5 ppb
T90 Response Factor
≤ 4 sec (typical)
**Range and MDQ (Minimum Detectable Quantity) are based on isobutylene.
Physical Characteristics
Weight
< 8 grams
Package Type
4P cell dimensional profile
Serviceable Parts
Lamp, detector cell, filters (2), cap, spacer
Lamp Life
Guaranteed 6000 hours (typical life significantly longer)
Onboard Filters
Removes liquids / particles
Warranty Period
18 months
Electrical Characteristics
Supply Voltage
3.2 V to 5.5 V (input voltage regulator included)
Current
24 mA to 36 mA
Power Consumption
80 mW to 200 mW (dependent upon supply voltage) 
Output Signal 
0.045 V to 2.5 V linear
Operating Specifications
Temperature Range
-20 °C to 60 °C (-4 °F to 140 °F)
Relative Humidity Range
0 to 90% non-condensing
Humidity Response 
≤ 1% @ 90% relative humidity
Humidity Quenching Effect
≤ 15% @ 90% relative humidity
Accuracy
± 3% of reading, with constant temperature and pressure


If you cannot find what you want, you can entrust ISweek to source for you. Just click:

Sourcing Service