Heat Flux Sensor - gSKIN-XM

• Ultra-high resolution of thermal energies and temperature differences.
• Low invasiveness and thickness.
• Models with connector compatible with all gSKIN®DLOG Data Loggers.
• All sensors with conductive heat flux calibration cohering to ISO 8301.

  • Model Number: gSKIN-XM
  • Data Sheet: Date Sheet File
Quantity:
- +
  • Heat Flux Sensor
  • Heat Flux Sensor
  • Heat Flux Sensor
Mouse over image to zoom

Product Specification

1.Heat Flux Sensor - gSKIN-XM Descriprion:

     Small heat flux sensor for performing reliable measurements under controlled conditions. Available both without and with connector.

 

2.Heat Flux Sensor - gSKIN-XM Features:

• Ultra-high resolution of thermal energies and temperature differences.
• Low invasiveness and thickness.
• Models with connector compatible with all gSKIN®DLOG Data Loggers.
• All sensors with conductive heat flux calibration cohering to ISO 8301.

 

3.Heat Flux Sensor - gSKIN-XM Applications:

     R&D, thermal optimization, energy efficiency, industrial monitoring of thermal properties.

 

4.Heat Flux Sensor - gSKIN-XM Technical parameters:

Product


gSKIN XM


Detector Type


Thermoelectric


Surface Material (Sensing Area)


Aluminum


Sensing Dimensions (mm)


4.4 x 4.4 x 0.5


Heat Flux Range Min / Max [kW/m²]


-150 / 150


Noise Equivalent Heat Flux per area [W/m²] / absolute [μW]


0.340 / 6.6 (Experimentally evaluated values under optimal steady state conditions).


Heat Flux Resolution per area [W/m2] with gSKIN® DLOG / absolute [μW]


0.41 / 7.9 (Guaranteed minimum heat flux resolution using the gSKIN® DLOG-4219)


Temperature Difference Resolution [μK]


140


Response Time (0-95%) [s]


0.7


Min. /Avg. Sensitivity (S) [μV/(W/m²)]


1.5 (Only a reference point).


Max. Compressive Force when clamped [kgf]


< 2


Electrical Resistance [Ohm]


< 20


Thermal Conductivity [W/(m*K)]


1.1


Linearity with Power [±%]


1


Operating Temperature Range Min/Max [°C]


-50 / 150


Calibration Accuracy [±%]


3


Calibration Temperature Range Min/Max [°C]


-30 / 70 (Conductive heat flux calibration cohering to the ISO8301 standard with mean temperature of 22.5°C).


Flexprint Length (f) [cm]


5


Temperature Dependence of S [%/°C]


0.25


Homogeneity [±%]


1 (The sensitivity increases or decreases as the temperature goes above or below 22.5°C). Position dependent signal change across sensing area.


Cable Length (c) [cm]


100


 




If you cannot find what you want, you can entrust ISweek to source for you. Just click:

Sourcing Service